User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDCÿ5000:Universal Controller DATE: 26-Feb-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 37 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F 1.002 ASK+ X 1.003 DISP Install UUT in mainframe and apply power. 1.004 DISP Allow the UUT 1 hour warm-up. 1.005 DISP On UUT, select SETUP until INPUT 1 is displayed. 1.006 DISP Select FUNC key 1 times for the IN1 TYPE display. 1.007 OPBR Does the UUT display T TC H? 1.008 JMPF 1.011 1.009 CALL Honeywell UDC 5000 Type T øF - High 1.010 JMP 1.022 1.011 OPBR Does the UUT display T TC L? 1.012 JMPF 1.015 1.013 CALL Honeywell UDC 5000 Type T øF - Low 1.014 JMP 1.022 1.015 OPBR Does the UUT display 4 - 20mA? 1.016 JMPF 1.019 1.017 CALL Honeywell UDC 5000 4 - 20mA 1.018 JMP 1.022 1.019 OPBR Does the UUT display 100 PT 1.020 JMPF 1.022 1.021 CALL Honeywell UDC 5000 100 PT 1.022 HEAD {} 1.023 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.023 DISP {º THIS COMPLETES THE VERIFICATION º} 1.023 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.024 END